Note 1:The meter size display loops as shown below: 
① 
  25mm      40mm        50mm      80mm            100mm        ② 
 
②   
 1inch     1.5inch     2inch      3inch           4inch    ① 
 
Note 2: When the meter size is changed, range unit and count rate will be forcefully changed as 
described below. If necessary, change these parameters again. 
 
Range unit 
m/s 
Count rate 
If the count rate goes out of the setting range   
when the meter size is changed, the count rate   
will be forcefully set to zero. 
 
 
Note 3: The exciting frequency setting may become inappropriate for the set value when the 
meter size of the detector is changed. If the exciting frequency is the value shown 
below when the meter size is changed, the exciting frequency will be forcefully 
changed. 
Setting meter size   
(mm) (inch) 
Set exciting frequency
25 1 
400Hz 
40 ~ 80 
1.5 ~ 3 
200Hz 
100 4 
100Hz