8.2.2 Exciting Current Value 
 
The exciting current value can be checked/changed by the following procedures. 
Be sure to match the exciting current value with the value specified for the 
combined detector

Specifying any other value may cause an error.
 
 
Shown below is an example of changing the exciting current value from 0.1900A to 0.2150A. 
Switch operation 
Display example 
Description 
 
A:DETECTOR 
1 EXC CUR
 
2 SIZE 
3 EXC FREQ 
4 FLOW DIR 
5 EXIT 
 
Select "EX CUR" from the setting item 
selection menu. 
 
STEP1 
 
 
 
EXC CUR 
 
0.1900 
 
         A 
 
 
The currently set excitation current value 
(0.1900A in this example) appears. 
 
Then push              . 
 
* Pushing        returns you to the setting 
menu. 
STEP2 
 
 
 
EXC CUR 
 
0
.1900 
 
         A 
 
 
The switches at the bottom change. 
(                                        are shown.) 
At the same time, the cursor appears. 
(The digit indicated by the cursor is 
highlighted.) 
 
Then push              . 
STEP3 
 
 
 
EXC CUR 
 
0.
1
900 
 
         A 
 
 
You can continue to change the setting 
value.  Push              to  move  the  cursor  to 
the digit you want to change. (You can hold 
down the switch longer for continuous 
operation.) 
 
Then push                . 
STEP4 
 
 
 
EXC CUR 
 
0.
2
900 
 
         A 
 
 
You can continue to change the setting 
value. Pushing      increments the 
number of the digit the cursor is positioned. 
(You can hold down the switch longer for 
continuous operation.) 
STEP5 
 
 
 
EXC CUR 
 
0.2150 
 
         A 
 
 
Repeat this operation to change the value to 
0.2150A
. 
When the desired value is obtained, push 
to set the value temporarily. 
 
ESC
ESC 
ESC