GE 100 User Manual

Page of 35
17
16
 
Meter Activation and Battery change
 
 
1. Turn the 
Meter over. Press and slide 
battery cover to open.
GE100 Blood Glucose 
CAUTION
- Please follow the local regulation and discard a used battery properly.
4. The Meter will enter self-tesing mode automatically when replacing the battery (all symbols will show 
on the screen).
5. Press any button to enter the Setting Mode.
6.  You  must  set  the  time  and  date  when  replacing  the  battery.  See  Chapter  "Setting  Up  Your  Meter  - 
Setting the Time, Volume and Date". Test results are still stored in the memory.
3. Slide the battery cover 
back until it snaps into 
place.
2. Install the battery. Be sure to put 
battery in correct direction 
(polarity (+) facing up).
 Meter Activation and Battery change
Your GE100 Blood Glucose Meter comes with one CR2032, 3volt, battery installed. One new battery will 
provide power to perform about 1000 tests under normal use. Press the main button or insert a strip to 
activate your meter.
You can enter Setting Mode two ways.
1. Replace Battery
    After  removing  the  battery,  press  the  main  button  several  times  until  there  is  no  signal  on  screen, 
then follow the battery installation steps to replace battery. The GE100 Blood Glucose Meter performs 
a self-test. Press the main button to exit the self-test and enter Setting Mode.
2. With Battery Inserted
  Press the main button first to turn on the GE100 Blood Glucose Meter. Hold down the main button for 
7 Seconds. During this time the screen will go blank and then after hearing a beep it will turn on into 
Setting Mode. The display screen will show setting data.