GE 100 User Manual

Page of 35
47
46
Limitations 
- Store the GE100 Blood Glucose Test Strips in the original capped vial at temperatures between 39  to 
86  ( 4  to 30  ) and relative humidity below 90%. Do not freeze.
- GE100 Blood Glucose Test Strips are designed for using with capillary whole blood samples. Do not use 
serum or plasma samples.
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (30048 meters) above sea 
level.
- Hematocrit below 30% may cause higher results, and hematocrit above 60% may cause lower results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- GE100 Blood Glucose Monitoring System has not been validated for use on neonates.
- Inaccurate test results may be obtained by abnormal concentrations of 
Uric acid ≧ 10.0 mg/dL, L-Dopa ≧ 3 mg/dL, Dopamine ≧ 2 mg/dL, Ascorbic acid (Vitamin C)≧ 5 mg/dL
°F
°F
°C
°C
NOTE
- Acetaminophen, Ibuprofen, Methyldopa, Salicylic Acid, Tetracycline, Tolbutamide, Bilirubin- 
  unconjugated, Cholesterol, Creatinine, Triglyceride, Maltose, Xylose, Galactose and Lactose 
  (when occurring in normal blood or normal therapeutic concentrations) do not significantly 
  affect results. However, abnormally high concentrations in blood may cause inaccurate results.
- Sodium Fluoride/Potassium Oxalate at low glucose levels may result in inaccurate blood glucose 
reading.  Sodium  Fluoride/Potassium  Oxalate  should  not  be  used  as  an  anticoagulant  with  this 
device.
Specifications
Memory Capacity
50 - 104
10 - 40
°F
°C
 (
)
10 - 90%
Turns  off  automatically  2  minutes  after  last  user 
action or press the "        " button for 3 seconds.
Measurement Technology
Measurement Calibration
Sample
Minimum Sample Volume
Measuring Range
Test Time
1.0 microliters
20-600 mg/dL
5 seconds
Oxidase Electrochemical Sensor
Plasma
Capillary whole blood
Operating Temperature
Operating Relative Humidity
Power Saving 
500 blood glucose test results with date and time