Hitachi Microscope & Magnifier S-3400N User Manual

Page of 466
3.10  Image Quality 
3 - 158 
3.10.2    Probe Current Setting and Image Quality 
 
On the ELECTORION BEAM block, image quality can be adjusted by specifying an appropriate 
probe current level.    An appropriate setting should be selected in reference to the table below. 
The default probe current is set to 50.    The lower the probe current, the higher the resolution at 
the expense of decreased probe current.    For low-magnification observation, the probe current 
should be set high, and for high-magnification observation, it should be set low. 
 
Table 3.10-2    Probe Current and Image Quality 
 
Probe Current Setting 
 
Irradiation current 
(A) 
Low High 
Resolution 
High Low 
Secondary electron signal 
Low High 
Image roughness 
High Low 
Charge-up 
※1 
Low High 
 
※1:  For non-conducting specimens 
 
See 3.4.1 Setting Parameters for the Electron Optical System. 
 
 
3.10.3    Objective Lens Movable Aperture and Image Quality 
 
The diameter of the objective lens movable aperture can be selected from four options. The 
table below shows the relationship between aperture diameters on one hand and the specimen 
current, the focal depth, and applications, on the other hand. Resolution is set so that it is the 
highest when the hole diameter is 30 µm.    The probe current can be adjusted by using the 
Probe Current menu on the Electron BEAM block. 
The surface condition of the objective lens movable aperture can have a significant impact on 
image quality, especially at low accelerating voltage levels. Maintenance services on the 
objective lens movable aperture should be performed periodically.    Also critical is axial 
alignment for the electron optical system. 
 
 
0              50              100