Toshiba lf516 User Manual

Page of 157
 
 
 92 − 
 
6 F 8 A 0 9 3 4  
 
Switch operation 
Display example 
Description 
STEP3 
 
 
 
H VAL 
 
  +10
3
.0 
 
       % 
 
 
 
Push                to  move  the  cursor  to  the 
digit you want to change and push
to change the number of the digit. 
 
 
 
STEP4(=END) 
 
 
 
H VAL 
 
  +103.0 
         
 
SET OK?
 
 
 
When you push       , a message appears 
to confirm whether the setting is OK or not. 
If OK, push       . If you want to redo the 
setting, push       . 
 
Pushing      cancels the setting operation 
and exits the setting screen. 
Note: If you try to set a value outside of the range 
−10% to +110%, LOW OVER SPEC or 
HIGH OVER SPEC
 error appears and the value returns to the previous value. Set a 
value once again. 
 
 
 
8.2.24  Mag-Prover
 
-Self Diagnosis ON/OFF Setting 
 
You can select on/off setting for Mag-Prover’s self-diagnosis function. 
If the self-diagnosis function is set to OFF, no error message is displayed even if any of the errors listed 
below occurs. 
 
• ROM error 
• RAM error 
• System parameter error 
• Excitation cable is not connected or its wiring is open 
• Excitation circuit failure 
• ADC circuit error 
• Totalizer data is destroyed 
 
Selection items 
Description 
OFF 
Self-diagnosis function is turned off. 
ON Self-diagnosis 
function is turned on. 
If this function is set to ON and an error occurs, an error message appears on the Display 2 measured 
value screen. If an error occurs, the measurement items specified for Display 2 screen cannot be 
displayed unless the error is removed. 