Toshiba lf516 User Manual

Page of 157
 
 
 104 − 
 
6 F 8 A 0 9 3 4  
9.2 Calibration Using Mag-Prover’s Built-In Signal Source 
 
9.2.1    0 % Flow Rate Calibration (Zero Calibration) 
 
Using Mag-Prover’s internal calibration circuit, 0% flow rate calibration (hereafter called zero 
calibration) can be performed. 
 
●Zero point check / calibration 
Switch operation 
Display example 
Description 
 
R:CAL 
1 CAL   0%
 
2 CAL  50% 
3 CAL 100% 
4 EXC DSPL 
5 EXIT 
 
 
Select "CAL 0" from the setting item 
selection menu. 
 
STEP1 
 
 
 
CAL   0% 
 
      0.1 
 
        % 
 
 
 
When the calibration screen is selected, the 
internal simulation circuit starts working and 
0%
 value using the internal simulation 
signal appears. 
Then push and hold       longer. 
* Pushing       returns you to the setting 
menu. 
STEP2 
 
 
 
Longer 
ADJ READY 
 
      0.1 
 
        % 
        
 
 
The title of the screen changes to “ADJ 
READY
” and the converter is is ready for 
calibration. 
 
Pushing       returns you to the previous 
screen. 
STEP3 
 
 
 
NOW 0% 
ADJUSTING 
 
          
        
 
 
Push       to start calibration for 0% flow 
rate. 
 
STEP4(=END) 
 
 
 
CAL   0% 
 
      0.0 
 
        % 
 
 
 
It takes several seconds to perform 
calibration for 0% flow rate and the 
simulated value of 0% after calibration 
appears. 
 
Pushing      returns you to the setting 
menu. 
Note 1:  To perform calibration, push and hold       longer. 
 
Note 2:  To cancel the adjustment when ADJ READY is displayed, push       .  
The screen returns to the zero display using the simulation input. 
 