6 F 8 A 0 9 3 4  
 
Switch operation 
Display example 
Description 
STEP6 
 
 
 
R2 
 
   
1
36.000 
 
     L/min 
 
 
 
Push       to change the number of the 
digit. 
Push              to move the digit. 
 
STEP7(=END) 
 
 
 
R2 
 
   100.000 
 
     L/min 
SET OK?
 
 
 
When you push       , a message appears 
to confirm whether the setting is OK or not. 
If OK, push       . If you want to redo the 
setting, push       . 
 
Pushing      cancels the setting operation 
and exits the setting screen. 
 
Note: Unites of the measuring unit changes as shown below: 
 
m
3
          L          mL          bbl          gal          m 
    CUSTOM            qt              pt                ft 
 
/s                  /min                  /h                /d 
  (Time unit) 
 
However, the following first and second unit combinations cannot be selected:   
m/min
, m/h, m/d, ft/min, ft/h, ft/d 
In the case of custom unit, time unit is not displayed. 
In the case of custom unit, character string set in 8.2.8 “Custom Unit Setting” is displayed.   
Identification character “*” showing the custom unit is displayed at the head of custom unit. 
 
● Changing the hysteresis 
The hysteresis used for multi-range switching is set to 3% (with respect to Range 1) when the 
flowmeter is shipped from the factory, unless otherwise specified. 
 
The following is an example to change the hysteresis from 3% to 5%. 
Switch operation 
Display example 
Description 
STEP1 
 
 
 
R HYS 
 
      03.0 
 
         % 
 
 
 
Select "R HYS" from the setting item 
selection menu. 
 
The currently set hysteresis (3.0% in this 
example) appears. 
 
Then push              . 
 
STEP2 
 
 
 
R HYS 
 
      
0
3.0 
 
         % 
 
 
 
The switches at the bottom change. 
(                                      are shown.) 
 
At the same time, the cursor appears. 
 
 
 
N O
O K
ESC 
O K
N O
ESC
First unit 
Second unit 