13.2.2 Flow rate indication is not correct. 
 
                                                                                                                           
                             
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Is the flow range correctly 
set?
 
Set correctly. Refer to   
Chapter 8, “Configuration 
Parameter Setting.”
 
NO 
YES 
Is zero point correctly set? 
Perform
 
the
 
zero
 
adjustment. 
Refer to 6, “Operation” 
NO 
YES 
Is
 
the
 
excitation
 
current
 
value as stated on the 
flow direction tag?
 
Set correctly. Refer to 8, 
“Configuration parameter 
setting”
 
NO 
YES 
Is
 
the
 
inside
 
wall
 
of
 
detector 
pipe contaminated? 
Clean the inside wall of the 
detector pipe.
 
YES 
NO 
Is
 
the
 
output
 
load
 
resistance 
as specified? 
Design the output circuit so that
 
the
 
output
 
load resistance is less 
than 750 
Ω
. 
NO 
YES 
Calculate as shown on the left. 
Are there two load   
resistors connected to the 
output in parallel? 
Connect the two load resistors 
in series, if necessary.
 
YES 
NO 
NO 
YES 
Contact your nearest Toshiba 
representative.
 
START 
Is accuracy calculated as follows? 
(Measured flow rate)-(Actual flow rate)
 
Actual flow rate
 
×
100%