22 Specifications
English  202052A
Check the 3M
TM 
SpotOn
TM
 Temperature Monitoring System website to ensure you have the most recent version of this document. 
www.spotontemperature.com reorder #202052A
Electrical 
characteristics
External power supply
100-240 VAC, 50-60 Hz 
Output; 5 VDC 
Class II, double insulated medical grade
Maximum heating power
2 W
Temperature 
characteristics
Over-temperature cutoff
43°C - Skin thermistor reading
44.5°C - Heater thermistor reading
Performance 
characteristics
Equilibration time
3 minutes
Time response
Approximately three minutes
Measurement locations
Patient’s lateral forehead above the orbital 
ridge
Measurement range
25°C - 43°C
Measurement readout
Celsius or Fahrenheit
Leakage current
CF classification. Meets leakage current 
requirements in accordance with UL 60601-1 
and EN60601-1.
Environmental 
conditions
Ambient temperature range
10°C to 40°C (50°F to 104°F)
Storage and transport temperature 
range
-20°C to 60°C (-4°F to 140°F)
Store all components at room temperature 
and in a dry place when not in use.
Operating humidity
10 to 75% RH, noncondensing
Atmospheric pressure range
80 kPa to 106 kPa (12 PSI to 15 PSI)
Oxygen environment
The sensor only is suitable for use in the 
presence of an oxygen enriched environment.