SITRANS LR 460 (HART) – INSTRUCTION MANUAL 
7ML19985JM01
mmmmm
E
: H
A
R
Co
mmu
n
ica
ti
o
n
s
Appendix E: HART Communications 
Highway Addressable Remote Transducer, HART, is an industrial protocol that is 
superimposed on the 4-20 mA signal. It is an open standard, and full details about HART 
can be obtained from the HART Communication Foundation at 
www.hartcomm.org
SITRANS LR 460 can be configured over the HART network using either the HART 
Communicator 275 by Fisher-Rosemount, or a software package. There are a number of 
different software packages available. The recommended software package is the 
SIMATIC Process Device Manager (PDM) by Siemens.
HART Device Description (DD) 
SITRANS LR 460 cannot be set up using a generic DD. The configuration software needs 
the HART Device Description for SITRANS LR 460.
You can download the HART DD for SITRANS LR 460 from the product page of our 
website. Go to: https://pia.khe.siemens.com/index.asp?Nr=14655 and click on Downloads. 
Older versions of the library will have to be updated in order to use all the features of 
SITRANS LR 460.
SIMATIC Process Device Manager (PDM)
This software package is designed to permit easy configuration, monitoring, and 
troubleshooting of HART devices. The HART DD for SITRANS LR 460 was written with 
SIMATIC PDM in mind and has been extensively tested with this software.
For more information see 
HART modem interface for SIMATIC PDM
To interface SIMATIC PDM with a HART device you will need a HART modem. Two 
options are available.
HART Modem model
Part Number
HART modem/RS-232 (for use with a PC and SIMATIC PDM)
7MF4997-1DA
HART modem/USB (for use with a PC and SIMATIC PDM)
7MF4997-1DB