SITRANS LR 460 – INSTRUCTION MANUAL 
Page B-77
mmmmm
B
: T
roub
le
shoot
in
g
Appendix B: Troubleshooting
1.
Check the following:
• There is power at the instrument.
• The LCD shows the relevant data.
• Check whether any fault codes are being displayed (see 
on page 80 for a detailed list).
• The device can be programmed using the handheld programmer.
2.
Verify that the wiring connections are correct.
3.
Check the HART (or PROFIBUS) address and make sure all devices are at unique 
HART (or PROFIBUS) addresses.
4.
If you continue to experience problems, go to our website at:
www.siemens.com/processautomation, and check the FAQs for SITRANS LR 460, or 
contact your Siemens Milltronics representative.
Troubleshooting LR 460
Symptom
Possible cause
Remedy
LR 460 reads 
100% 
continuously
False echo near antenna.
Move LR 460 to better location.
Rotate LR 460 to change polarization 
field.
Shorten vessel nozzle to allow antenna 
to project into vessel.
Aim LR 460 away from false echo.
Antenna buildup.
Clean antenna.
Install PTFE dust cover.
Use Air Purge option.
Reading is 
frozen
Antenna buildup.
Clean antenna.
Install PTFE dust cover.
Use Air Purge option.
Low dK and low bulk 
density material.
Lower echo confidence threshold.